Optimal design of a self shielded magnetic resonance imaging magnet
M. Souza, C. Vidigal, J. Taquin and M. SauzadeInstitut d'Electronique Fondamentale, Université Paris XI, Bât. 220, 91405 Orsay Cedex, France
(Received 26 April 1993, accepted $1^{er}$ September 1993)
Abstract
This paper describes an optimal design of a highly homogeneous resistive coil
system for Magnetic Resonance Imaging (MRI). Magnetic material is used to improve
field uniformity at the central region and to shield the magnet. The influence
of magnetic material is calculated by using a code based on the solution of scalar
and vector potentials equations. The obtained result is an axisymmetric coil
configuration enveloped by iron whose optimization was made by fixing one of
the criteria: the weak stray field near the magnet. It presents a great accessibility
to the homogeneous area and satisfies the bore's required dimensions. Dimensions
and field charts are given.
Résumé
Cet article présente une configuration optimale d'un aimant résistif très homogène
pour I.R.M. Un matériau magnétique permet à la fois de parfaire l'homogénéité au
centre et de réaliser le blindage de l'ensemble. les calculs ont été effectués à
l'aide d'un code utilisant les notions de potentiels scalaire et vecteur. Le résultat
obtenu concerne un système de révolution avec des cylindres de fer entourant les bobines.
Conçu pour un imageur corps entier, il présente une grande accessibilité du fait de
son faible encombrement en longueur. Les dimensions des différents éléments et les
cartes de champs sont présentées.
© Les Editions de Physique 1993